- Size Range: 0.01μm-1000μm
- Advanced light path design
- Built-in intelligent liquid dispersion units
- Three dimension automatic light path alignment system
- Users can freely select between manual and automatic modes
Wet Laser Particle Size Analyzer, PSA-LA2600A
説明
Application
Abrasives, adhesives, agrochemicals, barite, batteries, bentonite, boron carbide, brucite, bubble, calcite, calcium carbonate, carbon black, catalysts, cement, ceramics, chemicals, clay, coal, coatings, corundum, cosmetics, diamond powder, dolomite, diatomite, emulsions, environmental protection, explosives, ferrite, flour, fluorescent materials , fluorite, food and beverages, food additives, graphite, grinding , ink, kaolin, medicine, metal powder, mica, milling, minerals, oxides, paints, paper, petrochemicals, pharmaceuticals, pigments, gypsum, plastics, polymers, quartz, refractory materials, resins, silica, slurry, soil sediments, starch, sulfur, synthetic materials, talc, Toners, tourmaline, wollastonite, zeolite, zirconium silicate, etc.
Features
- Advanced Light Path Design
The converging light Fourier transform technology releases scattered light with large scattering angles from the restriction of the Fourier lens aperture, shortens the focal length, and improves the resolution of the instrument.
The ring-shaped multi-element silicon photo-diode ensures that the light signals of all particles are collected, greatly improving the resolution.
- Built-in Intelligent Liquid Dispersion Units
The stirring mechanism, ultrasonic dispersing unit, and sample circulation pipes are carefully aligned and integrated internally.
This built-in design effectively prevents issues such as inhomogeneous dispersion and the sedimentation of large particles, which commonly occur in external systems due to their excessively long sample circulation pipes. Consequently, the sample is uniformly and sufficiently dispersed.
- Unconstrained Free Fitting Analysis Techniques
The particle analysis software uses an unconstrained data fitting analysis technique to obtain the the real particle size distribution data, which is particularly important for researchers.
- Modern Measurement Control(Intelligent Sop Operation)
Users can perform all measurement procedures by simply operating on the PC and obtain ideal results in a very short time.
- User-friendly Operation
Users can freely select between manual and automatic modes to conduct measurements based on the sample’s characteristics. Under certain conditions (e.g., when the sample has unknown features or special measurement requirements exist), users may first perform a test measurement in manual mode. Once they have a clear understanding of the sample’s characteristics and measurement conditions, they can measure the samples in the automatic mode.
- Three Dimension Automatic Light Path Alignment System
A high-precision four-phase hybrid stepping motor automatically aligns and dynamically adjusts the optical path with a precision of up to 0.1 µm. This frees users from manual adjustments, improving the accuracy and stability of measurement results.
- Quick Measurements
In “Automatic” mode, all operational procedures are performed automatically, including water supply, stirring, circulation, background testing, sample testing, analysis, draining, and cleaning. This significantly reduces measurement time, with the entire process taking less than 2 minutes.
- Data Analysis
Data errors are rejected, and the measurement results are processed automatically. This eliminates the need for manual data processing and ensures standardized output.
Specification
Model | PSA-LA2600A | |
Measuring Principle | Laser light scattering | |
Analysis | Mie and Fraunhofer scattering | |
Size Range | 0.01μm-1000μm | |
Detectors | 87pcs./Log-spaced array/High sensitive/Photoelectric probe | |
Accuracy Error | <0.5% (Deviation of D50 on national standard sample) | |
Repeatability Error | <0.5% (Deviation of D50 on national standard sample) | |
Laser | High performance Fiber Laser λ= 639nm, p>2mW
Auxiliary laser λ=405nm, p>2mW |
|
Wet Dispersion | Ultrasonic | Frequency:40kHz, Power:60W, Time:≥1s |
Agitator | Revolutions Speed: 0-3000 rpm (Adjustable) | |
Circulation | Centrifugal pump, Rated flow:3-8L/min, Rated power:15W | |
Sample tank | Volume: 450mL | |
Operation Mode | Manual and Full automatic | |
Optics | Red light source/Reverse Fourier (convergent beam) | |
Optical Alignment System | Full automatic optical path alignment system | |
Software Function | Analysis mode | Free distribution, R-R Distribution, Logarithm normal distribution, Mesh number classification etc. |
Statistic method | Volume distribution, Quantity distribution | |
Statistic comparison | Several testing results of samples,
Different batches of samples testing result, Samples before and after processing, Test result of samples in different time |
|
User-defined analysis | Figure out percentage according to the particle size
Figure out particle size according to the percentage Figure out percentage according to the particle size range Meet demands of representation of particle test in different industries |
|
Test report | Word, Excel, JPG, Text and etc. | |
Multiple-language support | Multiple language support | |
Intelligent operation | Automatically control water inflow, dispersion, test and analysis;
Better repeatability after removing human factors; |
|
Testing Speed | <2min/time (including all the procedures), fastest measuring time<10s | |
Running Temperature | 15℃~35℃ | |
External Dimension (L*W*H) | 850*390*450mm | |
Net Weight | 40kg |
Get Social